Home > product > component

 

Insulation Tester is generally performed with DC500 V or DC1000 V, but there are no regulations whichever voltage should be used on test. Only, in general, low-voltage product(less than 150 V) is tested with 500 V, high-voltage products with 1000 V. In case of converting the leakage current at this time into resistance value ,which is higher than any standard value(when insulation resistance is high),the product is decided to be Good.

 

Product Model : KT-1000i
Model Output voltage Range of resistance measurement Function
KT-1000i Dimensions
W420, H140, L440
2 range of
DC 1000 V and DC 500 V
when DC1000 V : 0~2000
when DC500 V : 0~1000
Auto decision type

Merits of KAST Products
High accuracy
Measurement error is low due to three point adjustment type.
Auto decision type
If insulation efficiency is better than standard set value(if the tested value is high), it is
decided to be Good, and if insulation efficiency is low, it is Bad.
Decision speed is quick
It is convenient for a quick mass test due to measurement time 0.5 sec.
Easy to use
In case of setting standard value, only set a pointer to the needed value,
with pushing the button for limit set.
Strong
Strong to Impulse and noise because it is designed on the basis of many know-hows
through on-the-spot experiences.